Name: | Description: | Size: | Format: | |
---|---|---|---|---|
298.79 KB | Adobe PDF |
Advisor(s)
Abstract(s)
ABSTRACT: Hydrogenated amorphous carbon (a-C:H) top layer was deposited onto 316L medical grade stainless steel by reactive magnetron sputtering with two types of interlayers, based on Cr and Ti. The goal was to evaluate the effect of the adhesion-promoting layers on the overall coating/substrate system. After the immersion test in acidic artificial saliva, no surface signs of corrosion were detected. However, the Cr-based interlayer enhanced the metallic ions release, whereas the Ti-based one seems to improve the corrosion resistance. Electrochemical Impedance Spectroscopy was also used to evaluate galvanic corrosion evolution at the adhesion interfaces.
Description
Keywords
DLC coatings Metal release Impedance spectroscopy Orthodontics
Citation
Fróis, A... [et.al.] - Corrosion behaviour in artificial saliva of medical grade SS316L sputter-coated with DLC films: the effect of adhesion layers. In: Livro de Resumos : 8as Jornadas de Corrosão e Proteção de Materiais, Fórum Tecnológico, Lisboa, 24 novembro, 2022, p. 59-61
Publisher
LNEG - Laboratório Nacional de Energia e Geologia